Accelerometer Calibration System 3120BK

Model 3120BK is a back to back accelerometer calibration system for low frequency, low g level measurements.

Applications in which it is used include:

  • Back to back accelerometer calibration
  • Low Frequency, Low g level measurements
  • NIST traceable calibrations (when used with appropriate electrodynamic shaker, power supply and control system)

The Dytran 3120BK is an accelerometer calibration system.

Firstly, this model has 10 mV/g sensitivity and mounts to a shaker head via a 10-32 tapped hole.

Secondly, a vibration lab with an electrodynamic shaker and control system can use the model to accomplish in house NIST traceable calibrations. It offers back to back comparisons with a double ended calibration standard accelerometer.

This is achieved by mounting the test accelerometer directly to the mounting surface of the calibration standard (model 3120B). Then driving the coupled pair with an electrodynamic at various frequencies and amplitudes. At each test frequency the amplitude is set and the corresponding output read from the test accelerometer. The system includes model 4119B power supply unit, 3120B accelerometer as well as cables and accessories.

Other features include:

  • Stainless Steel
  • Calibration Standard Accelerometer
  • IEPE

 

Contact us for more information.

 

Accelerometer Calibration System 3120BK

Model Specifications
3120BK 1.02 mV/m/s² sensitivity, 4905 m/s² range, 10–32 radial connector, 10–32 tapped holes top and bottom, 85 grams,–51 to +93°C Operation

Reviews

There are no reviews yet.

Be the first to review “Accelerometer Calibration System 3120BK”

Your email address will not be published.

3120BLK